Model 405 Weighing Scale
User Instructions
1.0 General Information
Thank you for purchasing the Model 405 scale. Please read all operating
instructions carefully before using and note the following points:
Avoid lengthy exposure to extreme heat or cold. The scale works
best when operated at normal room temperature. Always allow the
scale to acclimate to a normal room temperature before use.
Allow sufficient warm up time. Turn the scale on and wait a few
minutes to allow the internal components to stabilize before
calibrating or weighing.
The cleaner the environment the better. Dust, dirt, moisture,
vibration, air currents and proximity to other electronic equipment can
cause an adverse effect on the reliability and accuracy of your scale.
Only use on a stable, vibration free surface.
Handle with care. Gently apply all items to be weighed onto the
weighing platform. Although this scale is designed to be quite
durable, try to avoid rough treatment as this may permanently
damage the internal sensor.
Place the item to be weighed on the platform and after the stable
weight is displayed, remove the item immediately. This will aid in the
longevity and accuracy of this weighing instrument.
Do not operate near an in-use cell phone, radio, computer or other
electronic device as these devices emit radio frequency (RF) and
may cause unstable scale readings.
Level the scale by looking at the bubble level under the display and
adjusting the feet on the bottom of the scale.
The scale may be powered by an AC adapter or a 6-volt
rechargeable battery. When Lo.bAt is flashes on the display, the
battery is low. Please recharge battery immediately.
Vista de página 0
1 2 3 4 5 6 ... 11 12

Resumo do Conteúdo

Página 1 - User Instructions

1 Model 405 Weighing ScaleUser Instructions1.0 General InformationThank you for purchasing the Model 405 scale. Please read all operating instructions

Página 2 - 3.0 Key Functions

107.0 Setting the AUTO-OFF FunctionYou can set the amount of time the scale is inactive before it shuts down to save battery power. Follow these steps

Página 3

11Specifications1. Capacity and division:405-6: 6.000kg X0.001kg / 12.000lb x 0.002lb405-15: 15.000kg x 0.002kg / 30.000lb x 0.005lb2. Display: seven

Página 4 - 4.0 Scale Operation

123. Working temperature: 5°C- 35°C4. Power supply: 12Vdc, 600mA AC adapter, with positive center or 6V4AH rechargeable battery.Note: When the rechar

Página 5 - 5.0 Calibration

22.0 Getting startedCharge the battery overnight with the AC adapter before using the scale or use the AC adapter to power the unit. The AC adapter pl

Página 6 - Calibration

3Press the On/Off/Zero key for more than 5 seconds to turn the scale on or off.ZERO FunctionWhen the scale is on, press the On/Off/Zero key for less t

Página 7

44.0 Scale OperationTurning on the scale:Be sure the battery is charged or that the AC adapter is plugged in.Press the On/Off/Zero key on the front pa

Página 8

55.0 CalibrationThis scale comes with the calibration jumper on the PC board in the CALIBRATE position. The calibration instructions below are written

Página 9

66.0 Accessing the PC board and Locking/Unlocking CalibrationTo lock or unlock the calibration mode for the scale you must access the PC board and mov

Página 10 - 8.0 Screen Messages

74. Turn the scale over again and carefully remove the top case from the base. Figure 1.1 shows the current main board and the calibration jumper (SWC

Página 11 - Specifications

8Figure 1.2 Older version of the main PC boardThe jumper shown in Figure 1.2 is in the ‘calibration unlocked’ position. The scale can be calibrated wi

Página 12

9Figure 1.3 Oldest style main PC boardThe jumper shown in Figure 1.3 is inside the small red rectangle. It is shown in the ‘calibration locked’ positi